During a labor epidural: an 18-gauge 3-1/2 epidural tipped needle was advanced into the epidural space from a midline.A multiorifice catheter was inserted to less than 10cm and resistance was met, then rotated needle 90 degrees to the right and attempted to advance catheter and still met resistance.Catheter was then withdrawn at that time and the blue tip was not intact.Provider unsure if blue tip broke during insertion or was never there.Therapy start date: (b)(6) 2018.Therapy end date: (b)(6) 2018.Diagnosis or reason for use: labor epidural.
|