Model Number 6935M62
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Cut In Material (2454); Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/05/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the right ventricular (rv) lead exhibited high and fluctuating impedance on the high voltage coil.When the lead was explanted and tested with an analyzer noised was also noted.At the midpoint of the lead an insulation breach was intentionally created to facilitate the introduction of a new lead.The lead was successfully explanted and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.Analysis indicated that the right ventricular defibrillation coil developed a fracture due to flexing while in vivo.The overlay tubing and the outer insulation of the lead was extrinsically breached due to a cut.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7411889
MDR Text Key104857453
Report Number2649622-2018-02944
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/08/2017
Device Model Number6935M62
Device Catalogue Number6935M62
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-