• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 27AGFN-756
Device Problem Device Expiration Issue (1216)
Patient Problem Endocarditis (1834)
Event Date 10/23/2007
Event Type  Injury  
Manufacturer Narrative
An event of valve explant due to endocarditis was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.It should be noted that the reported incident date was after the product "use by"/"use before" date.
 
Event Description
On (b)(6) 2007, a 27mm sjm regent heart valve was implanted.On (b)(6) 2017, the valve was explanted due to endocarditis.During the investigation of the explant, internal manufacturing records showed that the explanted device was implanted 10 months after its use by date of (b)(6) 2007.Abbott requested the implanting md review hospital medical records and to confirm the serial number and use by date.After review, the md confirmed that the internal information was accurate and at the time of initial implant, the hospital only stocked single sizes of valves and the expired valve was the only 27mm valve available for implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7412451
MDR Text Key104878678
Report Number2648612-2018-00030
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/08/2007
Device Model Number27AGFN-756
Device Catalogue Number27AGFN-756
Device Lot Number0002636354
Other Device ID Number81646879
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/14/2002
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age66 YR
-
-