Catalog Number 1070250-23
Device Problems Inflation Problem (1310); Leak/Splash (1354); Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/15/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Concomitant medical products: guide wire: bmw universal ii.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that the procedure was to treat a moderately tortuous distal posterolateral branch that did not have any calcification and was 90% stenosed.Pre-dilatation was performed with a 2.0 x 12 mm traveler balloon dilatation catheter (bdc).An attempt was then made to deploy a 2.5 x 23 mm xience xpedition stent; however, the stent delivery balloon failed to inflate after several attempts and kept losing pressure when inflated.The device did not meet any resistance during advancement.The stent delivery system was then removed from the anatomy and saline was injected through the catheter, which revealed a leakage coming from the proximal shaft as there was a tear.Therefore, the device was replaced with a 2.5 x 18 mm xience xpedition stent to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was returned for analysis.The reported tear and leak were confirmed.The reported inflation issues were not able to be confirmed due to the condition the device was returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported tear; however, it was determined the reported leak and inflation issues appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7413093
MDR Text Key105516424
Report Number2024168-2018-02601
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/21/2021
Device Catalogue Number1070250-23
Device Lot Number7031341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
-
-