Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: the device was returned for analysis.Device analysis revealed that the catheter shaft was checked for damage.No damage was noticed on the catheter shaft.The tip showed a small hole approximately 3mm from the very end of the tip.The damage was caused by the guide wire sticking through the tip of the device.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
|