SUNRISE MEDICAL TECHNOLOGIAS JAY XTREME ACTIVE CUSHION; JAY CUSHION Back to Search Results
Model Number JAY XTREME ACTIVE
Device Problem Material Integrity Problem (2978)
Patient Problems Pain (1994); Skin Tears (2516)
Event Date 03/13/2018
Event Type  Injury  
Manufacturer Narrative
The product in question is the jay xtreme active cushion.The rubber logo is sewn into the side of the cushion cover that sits below the rim of the cushion.This type of logo is used on covers for cushions that are manufactured and distributed in the united states as well; which is why this incident is being reported.This is the first time an incident of this kind has been reported to sunrise medical from any location world-wide.Since this cushion was not returned to sunrise medical, our quality team evaluated a similar cushion and attempted to duplicate the same effect without success that the end user is allegedly claiming is being caused by her cushion.The end user also claims that she does have feeling to the "bell" and thighs of her legs, however, kept using the cushion after several doctor visits even though it was allegedly causing the sores and scrapes.The end user has removed the logo from the cushion cover and has had no recurrence of the complaint.Sunrise medical's quality department will be performing an in depth assessment and monitor for trends.Sunrise medical's legal counsel is addressing her concerns and her claim.There will be no further investigation as there is no risk to public health identified.
 
Event Description
The following complaint was received on (b)(6) 2018 from an end user in (b)(6): i have used your jay pillow with cover for 20 years and it has gone well.I got a new cover for the pillow and after about a month i got sores and skin scrapes on my left bell / thighs.I find out that when i moved from the bed to the chair, the bell and thighs scratch your new logo.Your new logo is made of stiff rubber, it destroys my skin and i have to spend days in the hospital.I am now full of ugly scars and it hurts very much.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAY XTREME ACTIVE CUSHION
Type of Device
JAY CUSHION
Manufacturer (Section D)
SUNRISE MEDICAL TECHNOLOGIAS
no. 110 parque industrial
misiones de las californias
tijuana, baja california 22425
MX  22425
Manufacturer Contact
gustavo zambrano
2842 business park ave.
fresno, CA 93727
5592942840
MDR Report Key7427404
MDR Text Key105371680
Report Number9616084-2018-00003
Device Sequence Number1
Product Code IMP
Combination Product (y/n)N
Reporter Country CodeDA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Patient
Type of Report Initial
Report Date 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberJAY XTREME ACTIVE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/13/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-