(b)(4).Device evaluated by mfr: returned product consisted of the watchman access system (was).The device was bloody.The hub, sheath, and tip were microscopically, tactile and visually inspected.Inspection revealed tip damage (delamination), and kinks in the sheath located 18.5 cm and 72 cm from the strain relief.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
Reportable based on analysis completed 20 march 2018.It was reported that there was difficulty advancing the access system.A left atrial appendage (laa) closure procedure was being performed.A watchman ® access system was attempted to be advanced into the laa, but it was unable to be advanced because of the patient's anatomy.It was removed from the patient's body.The procedure was stopped.There were no patient complications and the patient is stable.However, device analysis revealed inner liner delamination on the tip of the access system.
|