Catalog Number BI70002000 |
Device Problem
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medtronic representative went to site to test the equipment.Representative reported that the voltage was not coming out of the power conversion board.Representative replaced power conversion board and updated firmware to resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect power conversion board has been received by the manufacturer for evaluation but the results are not available yet.
|
|
Event Description
|
A site representative reported that during spinal fusion procedure, the pendant of the imaging system displayed "initialization please wait".Rebooting the image acquisition system (ias) and mobile view station (mvs) while the umbilical was disconnected and then reconnecting the image acquisition system (ias) and mobile view station (mvs) did not resolve the issue.The procedure was completed without the use of imaging.There was a delay of less than 1 hour.No known impact on patient outcome.
|
|
Manufacturer Narrative
|
The power conversion enclosure was returned to the manufacturer for evaluation.Testing found that the voltage input was at zero on the board.It was noted that the indicator test did not pass.
|
|