• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/26/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to site to test the equipment.Representative reported that the voltage was not coming out of the power conversion board.Representative replaced power conversion board and updated firmware to resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect power conversion board has been received by the manufacturer for evaluation but the results are not available yet.
 
Event Description
A site representative reported that during spinal fusion procedure, the pendant of the imaging system displayed "initialization please wait".Rebooting the image acquisition system (ias) and mobile view station (mvs) while the umbilical was disconnected and then reconnecting the image acquisition system (ias) and mobile view station (mvs) did not resolve the issue.The procedure was completed without the use of imaging.There was a delay of less than 1 hour.No known impact on patient outcome.
 
Manufacturer Narrative
The power conversion enclosure was returned to the manufacturer for evaluation.Testing found that the voltage input was at zero on the board.It was noted that the indicator test did not pass.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7430741
MDR Text Key105656185
Report Number1723170-2018-01634
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age69 YR
Patient Weight75
-
-