DEPUY SYNTHES PRODUCTS LLC 5.0 CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT
|
Back to Search Results |
|
Model Number SHORT |
Device Problems
Overheating of Device (1437); Smoking (1585); Device Inoperable (1663); Metal Shedding Debris (1804)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/23/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This is report 3 of 3 for the same event.It was reported that during an ear nose and throat procedure, it was observed that the motor device and two attachment devices had warmth building in the surgeons hand after several minutes of holding the drill.It was also reported that smoke and metal shavings were then witnessed after which time the drilling was ceased.It was not reported if there were any delays in the surgical procedure.A spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was not confirmed.It was further reported that the device failed for vibration due to bearings being worn and corrosion from improper cleaning.It was further determined that the device had a damaged nose tube that flared out at the end which is indicative of excessive side loading causing the cutter to flex and to come in contact with the nose tube.The assignable root cause was determined to be due to user error and improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|