BARD ACCESS SYSTEMS SITE~RITE 8 ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770500
Device Problem Poor Quality Image (1408)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will be evaluated.Results are expected soon.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
End users are reporting that sr 8 system's image is too dark causing them to be unable to view the vessels and the needle properly during procedures.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: during evaluation, the reported issue of a dark ultrasound image could not be reproduced.The scanner was evaluated and the image was found to be normal for a sr 8 scanner with the 32 mm probe.There were no issues identified with the equipment.Manufacturing records were reviewed and there were no events that could have contributed to the reported complaint.The sr8 was serviced, tested, and returned to the customer.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
End users are reporting that sr 8 system's image is too dark causing them to be unable to view the vessels and the needle properly during procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
sydney freckleton
605 n. 5600 w.
salt lake city, UT 84116
8015225992
MDR Report Key7433122
MDR Text Key105679624
Report Number3006260740-2018-00716
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00801741096891
UDI-Public(01)00801741096891
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770500
Device Catalogue Number9770500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2018
Is the Reporter a Health Professional? Yes
Event Location Hospital
Date Manufacturer Received08/07/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-