Catalog Number 8065977763 |
Device Problem
Torn Material (3024)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: the device was not returned for evaluation.Complaint history and monarch product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.Additional information has been requested.(b)(4).
|
|
Event Description
|
A customer reported cartridge split during an intraocular lens (iol) implant procedure.Additional information was requested.
|
|
Manufacturer Narrative
|
Additional information provided.Cartridge product history records were reviewed and documentation indicates the product met release criteria.The lens product history record could not be reviewed because facility did not provide a lot number or any identification traceable to the manufacturing documentation.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
The cartridge was returned in a bag with the opened peel pouch.Only a small amount of solution was observed dried in the cartridge.The tip is heavily stressed.The anterior right side is cracked, starting prior to the parting line.The crack extended into an aneurysm that has torn and split through the tip end.The cartridge wings show evidence of being placed into a handpiece.The cartridge cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and the documentation indicated the product met release criteria.The customer indicated the use of an intraocular lens.The lens model has been qualified for use in the cartridge within the diopter range of 6.0 to 27.0.Without the associated lens diopter it cannot be determined if this is a qualified combination.Intraocular lens (iol) product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The customer indicated the use of a qualified handpiece and viscoelastics.The reported tip damage was observed.The root cause for the damage may be related to a failure to follow the directions for use (dfu).An inadequate amount of viscoelastic was observed.The dfu instructs to fill the cartridge with viscoelastic before loading the lens.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with viscoelastic, which may result in damage or delivery issues.The damage observed to the tip of the used cartridge typically occurs if the lens is not positioned correctly/folded correctly for advancement; if there is a lack of viscoelastic between the lens and the cartridge lumen; and/or if the handpiece plunger is not positioned correctly at the trailing optic edge.This can allow the lens to fold around the plunger tip making it too large to correctly advance through the narrow tip of the cartridge causing damage.(b)(4).
|
|
Search Alerts/Recalls
|