Two photographic images of the distal end of the blue introducer sheath were provided for evaluation.In both of the images a tear in the distal tip is present and the tear appears to have formed a flap in the blue introducer sheath.From the two images it appears that all components of the blue introducer sheath are accounted for.The device, including the dispenser gun, blue introducer and a 91cm catheter was returned for evaluation.Evaluation of the dispenser gun confirmed no abnormalities and the gun was fully functional.During evaluation of the catheter, biological material which had occluded inside the tubing was observed at the distal end of the white catheter.During visual inspection of the blue introducer sheath, a tear in distal tip was observed which appears to have formed a flap in the blue introducer sheath.If information is provided in the future, a supplemental report will be issued.
|