Model Number N/A |
Device Problems
Device Packaging Compromised (2916); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Information (3190)
|
Event Date 03/19/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).Reporter had indicated that product will be returned.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that foreign substances were inside of the sterile packages.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned package determined that there was a blue-colored stain present on the sterile packaging.Device history record was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to manufacturing deficiency.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|