Catalog Number BI70002000
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.Representative reported that when looked at logs, collision zones warning was noticed.Some movements buttons do not respond until out of collision zone.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Event Description
Medtronic received information that during spinal fusion procedure, the software of the imaging system became unresponsive.Rebooting the software resolved the issue.The procedure was completed with the use of imaging.There was a delay of 10-15 minutes.No known impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7443956
MDR Text Key106147116
Report Number1723170-2018-01682
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse