Model Number C0936731
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/29/2018
Event Type  malfunction  
Manufacturer Narrative
Samples received: 1 open pouch.Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed (b)(4) units in the market.There are no units in stock in b.Braun surgical warehouse.We have received an open and used suture with the needle detached from the thread as can be seen in the enclosed picture.However, without closed samples a proper analysis cannot be performed.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Final conclusion: in spite of receiving a defective sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyze it.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Event Description
Country of complaint: (b)(6).Thread detached from needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 4/0 (1.5) 75CM DSMP19 (M)
Type of Device
SUTURES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP  08191
Manufacturer (Section G)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP   08191
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key7446941
MDR Text Key106187028
Report Number3003639970-2018-00212
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K990090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/29/2021
Device Model NumberC0936731
Device Catalogue NumberC0936731
Device Lot Number616175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2018
Distributor Facility Aware Date03/29/2018
Date Manufacturer Received03/29/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-