ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER LASER PROBES; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED Back to Search Results
Catalog Number 8065751593
Device Problem Sticking (1597)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The sample has been received and in-house evaluation is in progress.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that during surgery, on multiple occasions, the curved laser probe stuck when inserting inside the trocar.The probe was switched out on each case to complete surgery.There was no patient harm.
 
Manufacturer Narrative
Additional information provided.The 25 gauge illuminated flex curved laser probe lot number was received and a visual assessment of the returned sample revealed significant adhesive on the cannula tip.The 25 gauge illuminated flex curved laser probe was inserted into a 25g trocar.The sample had resistance when inserted into the 25g trocar, confirming the reported event.Additionally, the sample¿s cannula outer diameter (od) was measured.With an od min/max values of 0.0200 to 0.0205 inches, the sample¿s cannula od was measured to be 0.02035 to 0.02085, not meeting specifications.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to excess adhesive residue on the laser probe tips during manufacturing.An internal investigation has been opened.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LASER PROBES
Type of Device
INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7450397
MDR Text Key106384691
Report Number2028159-2018-00823
Device Sequence Number1
Product Code HQE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K946135
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number8065751593
Device Lot Number17024845X
Other Device ID Number380657515936
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-