Catalog Number 295600-002 |
Device Problems
Electrical /Electronic Property Problem (1198); Device Inoperable (1663); Inadequate Lighting (2957)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because this issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of onboard batteries.The freedom home ac power supply has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom home ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom home ac power supply was not working; the led light did not illuminate when inserted in the patient's freedom driver.The customer also reported that the patient was provided a backup freedom home ac power supply.
|
|
Manufacturer Narrative
|
Visual inspection of the ac power supply revealed a cracked white connector and crack in the housing of the power brick.The cause of the customer-reported issue of the led not illuminating was determined to be a malfunction of the internal printed circuit assembly (pca), where the l3 component had detached from the board.The root cause of this detachment cannot be confirmed, but it is likely the result of a drop or other rough handling, given the observations made during the visual inspection.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|
Search Alerts/Recalls
|