Catalog Number 295600-002
Device Problems Electrical /Electronic Property Problem (1198); Device Inoperable (1663); Inadequate Lighting (2957)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because this issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of onboard batteries.The freedom home ac power supply has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom home ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom home ac power supply was not working; the led light did not illuminate when inserted in the patient's freedom driver.The customer also reported that the patient was provided a backup freedom home ac power supply.
 
Manufacturer Narrative
Visual inspection of the ac power supply revealed a cracked white connector and crack in the housing of the power brick.The cause of the customer-reported issue of the led not illuminating was determined to be a malfunction of the internal printed circuit assembly (pca), where the l3 component had detached from the board.The root cause of this detachment cannot be confirmed, but it is likely the result of a drop or other rough handling, given the observations made during the visual inspection.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOME AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7453209
MDR Text Key106869071
Report Number3003761017-2018-00124
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Remedial Action Replace
Type of Report Initial,Followup
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295600-002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age53 YR
-
-