• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN SURGICAL SA HISTOACRYL BLUE TISSUE ADHESIVE 0.5ML; TISSUE ADHESIVES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA HISTOACRYL BLUE TISSUE ADHESIVE 0.5ML; TISSUE ADHESIVES Back to Search Results
Model Number 1050052
Device Problems Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Manufacturing site evaluation: evaluation on-going.Device not returned.
 
Event Description
Country of complaint: (b)(6).Leakage was found after opened the package.
 
Manufacturer Narrative
Samples received: 5 unopened and 2 opened pouches received: analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed in the market 4,210 units.There are no units in stock in b.Braun surgical's warehouse.We have received 5 unopened and 2 open pouches with an ampoule.The five closed samples have been opened and optically evaluated and the ampoules inside do not present leakage.The other two ampoules received from an open pouch have also been optically evaluated and a defect in the tip of the ampoule has been found.The leakage of the glue occurs at this point.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Final conclusion: taking into account that the results of the samples received do not fulfil b.Braun surgical specifications, we conclude that the complaint is confirmed by evidence of failure in the samples received.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HISTOACRYL BLUE TISSUE ADHESIVE 0.5ML
Type of Device
TISSUE ADHESIVES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key7454148
MDR Text Key106442842
Report Number3003639970-2018-00232
Device Sequence Number1
Product Code MPN
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1050052
Device Catalogue Number1050052
Device Lot Number217381N2
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/19/2018
Date Manufacturer Received03/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-