Model Number 595000-001 |
Device Problems
Detachment Of Device Component (1104); Device Slipped (1584); Component Misassembled (4004)
|
Patient Problems
No Patient Involvement (2645); No Information (3190)
|
Event Date 04/13/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed while the freedom driver was not supporting a patient.The freedom driver will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the vent cover on the freedom driver was loose and came off when the driver was pulled out of the shipping case.
|
|
Manufacturer Narrative
|
The customer-reported vent cover not secured was confirmed during incoming evaluation of the driver, where one screw on the right side of the exhaust fan cover was loose and the other was unscrewed.The cause of the unsecured exhaust fan cover was determined to be a workmanship issue in that the fastening screws that secure it to the housing were not properly inserted/ tightened during the service procedure.Despite the unsecured exhaust fan cover, the driver passed all functional testing.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|