MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1711K |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/08/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that the insulin pump's screen was flashing on and off and the pump was not reacting.Customer's blood glucose value was 5.8mmol/l.Insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Device received with a constant blank flashing white light on the display followed by an unexpected intermittent beep alarm due to vertically cracked lcd controller.Unable to perform the self test, sleep current measurement, active current measurement, and displacement test verify missing segments/partial display due to blank display.
|
|