SMITH & NEPHEW, INC. BIORAPTOR KNOTLESS SUTURE ANCHOR, HIP; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202397
Device Problem Bent (1059)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/23/2018
Event Type  malfunction  
Event Description
It was reported that the anchor bent inside the arthroscope.No significant delay or patient injury reported.
 
Manufacturer Narrative
Examination was not possible, as the device has not been returned.The investigation could not draw any conclusions about the reported event without the return of the device.
 
Manufacturer Narrative
One bioraptor knotless suture hip anchor was returned for evaluation.Visual assessment of the device shows it is complete.The anchor tip is bent and the inner grub screw has been advanced out of the proximal end of the anchor.The torque limiter was functionally tested and found to perform as intended.The damage to the anchor indicates the inner plug was advanced excessively counter clockwise during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIORAPTOR KNOTLESS SUTURE ANCHOR, HIP
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7463233
MDR Text Key106625563
Report Number1219602-2018-00492
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010651280
UDI-Public(01)03596010651280(17)220516(10)50664571
Combination Product (y/n)N
PMA/PMN Number
K093428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup,Followup
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/16/2022
Device Model Number72202397
Device Catalogue Number72202397
Device Lot Number50664571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Date Manufacturer Received06/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age83 YR
-
-