The field generator was returned for evaluation.The returned field generator was plugged into a known good axiem, but was unable to connect.Further investigation revealed that the lemo connector was partially unscrewed.The lemo connector was tightened, but the field generator was still unable to connect.All pins inside of the lemo connector visually appeared good.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
Medtronic received information that, while in a catheter placement procedure, the emitter on the navigation system was showing "localizer not connected." the representative had the site swap the system emitter with an emitter from another navigation system on-site and was able to track.There was a delay of less than an hour to the procedure due to the alleged issue.There was no reported impact on patient outcome.
|