Model Number 5392
Device Problems Connection Problem (2900); Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that one of the external pulse generator's cable connectors was broken.The generator has not been received into service.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis confirmed the stated reason for return of "one of the connectors of the cable is broken", the ventricular connector was broken.It was additionally noted that the hook assembly was broken.The patient connectors and the hook assembly were replaced.The device passed all tests with no visual or electrical anomalies.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the product was returned for service.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7464143
MDR Text Key106715181
Report Number3004593495-2018-00421
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received09/05/2018
Date Device Manufactured06/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-