Model Number M004RC64S0
Device Problems No Display/Image (1183); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/15/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: analysis of the returned device found the deployment shaft and magnetic sensor wires separated due to overload force.Electrical testing was performed and the device failed the test.The magnetic sensor resistance and inductance measurements were within specification.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion was unable to be determined.(b)(4).
 
Event Description
Reportable based on analysis completed april 10, 2018.It was reported that when the intellamap orion¿ mapping catheter was connected, an error message appeared indicating the magnetic sensor was either disconnected or broken.Signals were visible, but the catheter was not tracked or displayed on the rhythmia mapping system.Replacing the catheter resolved the issue.There were no patient adverse events.However, returned device analysis found the deployment shaft had separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7464236
MDR Text Key106634411
Report Number2134265-2018-03707
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K122461
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/13/2018
Device Model NumberM004RC64S0
Device Catalogue NumberRC64S
Device Lot Number21130129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MAPPING SYSTEM: BOSTON SCIENTIFIC RHYTHMIA
-
-