Catalog Number 1012446-15 |
Device Problems
Detachment Of Device Component (1104); Kinked (1339); Difficult To Position (1467)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/03/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
It was reported that the procedure was to treat a 90% stenosed, moderately tortuous, and heavily calcified de novo lesion in the mid left anterior descending artery.Following pre-dilatation with multiple balloons, a 2.25x15mm nc trek balloon dilatation catheter (bdc) was being reused; however, the bdc became stuck in the guiding catheter during advancement.The proximal shaft kinked and separated.The separated portion was simply withdrawn out of the guiding catheter and the procedure was successfully completed with another unknown device.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
Internal file number - (b)(4).Evaluation summary: visual inspection was performed.The reported shaft separation and kink were confirmed.The reported difficulty positioning the device could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that the reported difficulty positioning the device contributed to the reported kink and separation.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
|
Search Alerts/Recalls
|