MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problems Erratic or Intermittent Display (1182); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/03/2018
Event Type  Injury  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The imaging system then passed the system checkout and was found to be fully functional.The software was returned for analysis and investigation found that the reported event was related to a known event.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Event Description
Information was received regarding a navigation system issue that occurred during a cranial resection.It was reported that during the registration the screen went black and displayed an error code.The site rebooted the system with no resolution.The surgery was rescheduled.There was no reported impact on patient outcome.There was a reported delay to the procedure of less than 1 hour due to this issue.No further information provided.
 
Manufacturer Narrative
It was reported that the patient was under anesthesia when the reported issue occurred and the procedure was subsequently rescheduled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7466054
MDR Text Key106664627
Report Number1723170-2018-01768
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Catalogue Number9735665
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age50 YR
Patient Weight67
-
-