This report is being submitted as follow up no.1 to provide the udi no., the expiration date, device return date, and the completed investigation results.The actual device was returned for evaluation.Visual inspection upon receipt revealed no defects.The actual sample, after having been rinsed, was built into a circuit with tubes and primed with colored saline solution in accordance with the ifu for this product.There was no air remaining inside the actual sample.The blood phase was filled with saline solution.With the blood outlet port clamped, air pressure was applied to the blood phase from the blood inlet port.No leak was confirmed.The actual sample was then circulated with bovine blood at the back pressure of 200mmhg and at each blood flow rate of 0.5l/min., 1.0l/min., and 1.5l/min.During circulation at each flow rate, an air of 10ml was sent into the circulation in 30 seconds.No air came through the filter out of the oxygenator module.Functional testing was conducted.A tube was connected to the luer port on the oxygenator module of the actual sample and the pump on the cardioplegia side was let to run.It was found that air was pulled into the oxygenator module when the flow rate of the blood going into the oxygenator module was lower than that going out of the oxygenator module, or when the pump on the extracorporeal circulation side was not running.There is no evidence that this event was related to a device defect or malfunction.The investigation result verified that the actual sample was the normal product without any leak which could lead air to enter the oxygenator module, having the normal air removal performance.As a possible cause of this complaint, the circulation of the line on the cardioplegia side may have led the pressure inside the oxygenator module to be negative and air may have been pulled into the oxygenation module through the fibers.However, the exact cause of the reported event cannot be definitively determined based on the available information.A review of the manufacturing record and the product release decision control sheet of the involved product/lot# combination confirmed there were no indications of production-related anomalies or discrepancies in the inspection results.A search of the complaint file did not find any other report of this nature with the involved product/lot# combination.The ifu states: to prevent gaseous emboli from entering the blood phase, make sure that the arterial pump flow rate always exceeds the flow rate of the cardioplegia line.The blood flow rate of the cardioplegia line should not exceed 0.5 l/min.
|