Model Number 3660
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inaccurate elective replacement indicator advisory notice issued by abbott on 12 september 2017.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported ((b)(6)) the ¿replace generator soon¿ message was displayed on the patient controller indicating the generator was approaching its end of service.The ipg log was assessed and the ipg was not at a true eri, the elective replacement indicator message displayed was premature.The device has the appropriate level of battery voltage to provide therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7477685
MDR Text Key107080067
Report Number1627487-2018-04349
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date 05/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/07/2018
Device Model Number3660
Device Lot Number5553964
Other Device ID Number05415067020192
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/09/12/2017/001-C
Patient Sequence Number1
Patient Age86 YR
-
-