The patient's son stated they received questionable results from coaguchek xs meter serial (b)(4).They used strip lots 27216421 and 25863822 on both dates of testing.He could not specify which lot gave which results.The medwatches involved in this event include (b)(6).On (b)(6) 2018, within 30 minutes the results from two different fingers were 2.3 inr, >8.0 inr, and 1.6 inr.He stated the patient ate some brussel sprouts in between the tests.No changes were made in the warfarin dose due to the results.On (b)(6) 2018, the patient performed three tests within less than 10 minutes and the results were 2.3 inr, 1.8 inr and 1.9 inr.The patient's son reported the result of 2.3 inr.There was no allegation of an adverse event.There were no recent changes in health or other medications, no other blood thinners except warfarin and baby aspirin.The patient had no direct thrombin inhibitors, no lupus, no antiphospholipid antibodies, and no concerns with hematocrit levels.The patient had no signs of high inr.The therapeutic range was 2.0-3.0 inr.The suspect meter and strips were requested to be returned for investigation.Relevant retention test strips (lot 258638) were tested in comparison with the master lot coaguchek xs pt.For this purpose, two human blood samples from marcumar donors and two internal reference meters were used.No error messages occurred.Retention samples were acceptable and retention material performed as specified.
|