MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA PRESSUREWIRE¿ X, WIRELESS, RX 175CM; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA PRESSUREWIRE¿ X, WIRELESS, RX 175CM; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number C12059
Device Problem Disconnection (1171)
Patient Problem Thrombus (2101)
Event Date 04/16/2018
Event Type  Injury  
Event Description
A pressure wire x, wireless could not be equalized and lost signal when advancing past the lesion.The light remained green but no signal was reaching the quantien.A second device was used and could not equalize due to a drift.In the 20 minutes while attempting to do the ffr the patient had a stent thrombosis.This complication required the patient to receive balloon angioplasty and be admitted to the hospital.The patient required a repeat angiogram.The patient is stable now after the physician treated.Patient's information (e.G.Age, weight, gender, patient's initials) cannot be handled by abbott in absence of patient's written consent as required by the personal data protection national legislation.National legislation prevents the recording of such information.A written consent has not been obtained in this case; therefore, this information is not available.Please reference 3008452825-2018-00141 for the second device.
 
Manufacturer Narrative
The reported event of signal loss and an event of extended procedure was confirmed.The results of the investigation confirmed the pressurewire was unable to calibrate due to open circuits located at the sensor chip area.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported signal loss is consistent with the open circuits.Based on the received condition of the device and the information provided, the cause of the open circuits remains unknown.The pressurewire aeris instructions for use (ifu) states that potential complications which may be encountered during all catheterization procedures include but are not limited to: vessel dissection or occlusion, perforation, embolus, spasm, local and/or systemic infection, pneumothorax, congestive heart failure, myocardial infarction, hypotension, chest pain, renal insufficiency, serious arrhythmias or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSUREWIRE¿ X, WIRELESS, RX 175CM
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7480854
MDR Text Key107152268
Report Number3008452825-2018-00132
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K161171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Model NumberC12059
Device Catalogue NumberC12059
Device Lot Number6073348
Other Device ID Number05415067025715
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/31/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
-
-