BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M004EPMR9620K20 |
Device Problems
Bent (1059); No Display/Image (1183)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: visual inspection revealed that the distal end was bent to the right.The seal to ring 1 appeared to be compromised and body fluid was present along the distal edge of the electrode.There was also dried body fluid on the luer, handle, and shaft.Electrical testing revealed no shorts, however the magnetic sensor was found to be open in all curve configurations.X-ray showed a bent center support.Functional inspection, rf ablation testing, and dissection revealed no anomalies.The manufacturing batch record review confirmed the device met all material, assembly, and performance specifications.The investigation conclusion is design specification as the device problem was traced back to the design specifications.(b)(4).
|
|
Event Description
|
This event is reportable based on device analysis completed on 4/4/18.While in the right ventricle, during ablation of ventricular tachycardia, the intellanav mifi catheter disappeared from the screen; no error messages were displayed.The curve of the catheter also looked deformed on x-ray.The catheter was removed from the patient to visually check the curve and at this time, no directsense data was displayed.Disconnecting the catheter and restarting the study did not resolve the issue.The catheter was replaced in order to complete the procedure, and no patient complications occurred.Device analysis revealed a compromised seal on ring 1.
|
|
Search Alerts/Recalls
|
|
|