(b)(4).Device evaluated by mfr: analysis of returned product revealed the unit passed electrical and rf ablation testing, consequently, not confirming the reported complaint of "noisy electrogram"; however, the device had a kink at the distal section-approximately the distal section while in the neutral position-ring 1 seals were compromised; there was dried body fluid on the edge of the electrode.Bent center support and a steering wire partially detached from the center support were observed under x-ray between r1 and r2 at the distal section.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|