Catalog Number 305343
Device Problem Difficult to Open or Close (2921)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/13/2018
Event Type  malfunction  
Manufacturer Narrative
The manufacturing location for this product is flextronics.This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed.The (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.Device manufacture date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
 
Event Description
It was reported that the temporary lids of a bd¿ multi-use nestable sharp collector, 8 qt hinge cap didn¿t shut.There was no report of exposure, injury or medical intervention.
 
Manufacturer Narrative
Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 7261929.Medical device expiration date: unknown.Device manufacture date: unknown.Medical device lot #: 7225923.Medical device expiration date: unknown.Device manufacture date: unknown.Medical device lot #: 7272938.Medical device expiration date: unknown.Device manufacture date: unknown.Medical device lot #: 7230933.Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: bdj received 6 boxes.(2 boxes of lot 7261929, 2 boxes of lot 7225923, 1 box of lot 7272938 and 1 box of lot 7230933.) bdj checked half lids from each box and found 1 lid of lot7261929(cavity#1510-3/2011-1)opened immediately after it was shut.The rest could keep closed over 10min.According to this investigation this failure mode is already known since previous complaints were received for the same condition throughout 2017 reason why a capa record # (b)(4) was opened to perform the investigation and to implement corrective actions.It was not confirmed if the defective product reported in this complaint arose from a lot manufactured after of the implementation of the corrective actions due to the lot number was unknown.Corrective actions were full implemented for batches manufactured from 2018 onwards.Investigation conclusion: based on this investigation this failure mode was caused due worn out on the mold, the corrective actions were implemented and evaluated as effective.Root cause slot reference dime:nsion was found greater (loose) than drawing specification.Supplier corrective actions align steel as per drawing requirement.Create mold inserts to be replaced if needed, during preventive maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ MULTI-USE NESTABLE SHARP COLLECTOR, 8 QT HINGE CAP
Type of Device
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
MDR Report Key7484356
MDR Text Key107323273
Report Number2243072-2018-00223
Device Sequence Number1
Product Code FMI
UDI-Device Identifier50382903053433
UDI-Public50382903053433
Combination Product (y/n)N
PMA/PMN Number
K943134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number305343
Device Lot NumberUNKNOWN
Date Manufacturer Received04/13/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-