|
Model Number H7491213001J2 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/09/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr: visual inspection howed the upn characteristics of this device were checked and effectively, it is design to use together with the magnet device.The device has the distal tip kinked.The device has the body broken approximately at 47 cm from the proximal end; the another returned section of wire measures approximately 135 cm from the distal tip.Also, it seems to be the broken section was kinked previous to be broken, due to both ends are bent.The overall length could not be performed due to device condition (body broken).The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on 13april2017.It was reported that magnet would not hold wire.A luge guidewire was selected for a angioplasty stent intervention however during the procedure doctor had the balloon in place, he was in the process of exchanging the balloon over the wire for a stent.In the process of removing the balloon using the magnet, the wire came back from across the lesion, wire position was not retained.The doctor removed the balloon and then used a rapid exchange system thereafter.They did not have to switch out to another magnet, the doctor refused to use another magnet.No complications to the patient, post procedure patient was stable.However analysis reveled that the guidewire was fractured.
|
|
Search Alerts/Recalls
|
|
|