Catalog Number FS-OMNI-35
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical product: erbe electrosurgical generator.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved could not be reviewed because a lot number was not provided.Investigation conclusion: a corrective action has been initiated to reduce occurrences of stripping difficulty for fs-omni sphincterotomes.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all fusion omni-tome pre-loaded sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.Corrective action: a review of the complaint history was conducted.A corrective action has been initiated in an effort to reduce occurrences of this nature.This product is included in the scope of this corrective action.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion omni-tome pre-loaded sphincterotome.The wire did not split out of the catheter properly.It was really difficult.[there was a] "stripping issue" with wire [guide] and plastic [catheter] at the time they were ready to make a device exchange for a balloon [difficult exchange].The problem was that the wire was not splitting properly away from the sphincterotome catheter and the customer lost access and positioning [lost wire guide access] (subject of this report).The case was aborted as it was a very challenging patient and anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key7485121
MDR Text Key107993947
Report Number1037905-2018-00176
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K052051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFS-OMNI-35
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/05/2018
Event Location Hospital
Date Manufacturer Received04/09/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-