Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 04/04/2018
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint for the platform displaying an error message user advisory (ua) 16 (timeout moving to take-up position) was confirmed during initial functional testing.The root cause was determined to be a defective system processor board.The platform failed the initial functional testing due to error message user advisory (ua) 16 displayed when the platform was powered on.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial number (b)(4).
 
Event Description
During shift check, the autopulse platform ((b)(4)) displayed a user advisory (ua) 16 (timeout moving to take-up position) error message.The user was unable to resolve the issue.No patient was involved with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
noemi schambach
2000 ringwood ave,
san jose, CA 95131
4084192955
MDR Report Key7488734
MDR Text Key107594078
Report Number3010617000-2018-00514
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111065009
UDI-Public00849111065009
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-03
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-