The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: during evaluation, the reported issue of the ultrasound image lag could not be reproduced.The scanner was restarted 15 times and no image lag was noted.The review of the dhr, sub-assembly dhr, sub-assembly manufacturing, component records, manufacturing process changes, mrr/mrb, and internal rejects showed that the reported issue is not likely related to a manufacturing issue.See referenced complaint for additional reported failure investigation details.The sr8 was serviced, qc tested, and returned to the customer.
|