• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION, INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION, INC AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/06/2018
Event Type  malfunction  
Manufacturer Narrative
The reported complaint was confirmed during the archive review.Several ua 45 errors were observed, however, the faults were not reproducible during the functional testing of the platform.The archive review showed customer complaint on ua45 (drive shaft not at home position) in the archive data on (b)(6) 2018.There's no indication of device being power up in the archive on the date problem reported of (b)(6) 2018.The device passed functional testing.No faults observed.
 
Event Description
The autopulse platform displayed ua45 (not at "home" position after power-on/restart) error message during shift check and the crew cannot cleared the error.No known impact or consequence to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION, INC
2000 ringwood avenue
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION, INC
2000 ringwood avenue
san jose CA 95131
Manufacturer Contact
kim nguyen
2000 ringwood avenue
san jose, CA 95131
MDR Report Key7489578
MDR Text Key107593747
Report Number3010617000-2018-00523
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Emergency Medical Technician
Type of Report Initial
Report Date 05/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/06/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-