Product event summary: the controller (b)(4) was not returned for evaluation.Log files covering the event date were not available for evaluation.As a result, the reported "incomplete data download" and loss of power event could not be confirmed.Applicable risk documentation and experience with events of similar circumstances were considered; events involving a controller that was unable to communicate to a monitor can be attributed, but not limited to, a serial port connector malfunction and/or a marginal internal connection between the serial port connector and the printed circuit board (pcb).A possible root cause of loss of power can be attributed to disconnection of both power sources from the controller and/or to an intermittent disconnection on one more both power sources.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|