Catalog Number 762165
Device Problem Short Fill (1575)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/21/2017
Event Type  malfunction  
Manufacturer Narrative
Date returned to manufacturer: unknown.The investigation date has been used.(b)(6).Bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for 128149 with the incident lot was not observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.
 
Event Description
It was reported that paxgene® blood rna tube "rna-762165-underfill".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAXGENE® BLOOD RNA TUBE
Type of Device
RNA PREANALYTICAL SYSTEMS
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7490975
MDR Text Key107594956
Report Number9617032-2018-02365
Device Sequence Number1
Product Code NTW
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
DEN050003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2018
Device Catalogue Number762165
Device Lot Number6239926
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-