BAXTER HEALTHCARE - ENGLEWOOD 4000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938742
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem No Patient Involvement (2645)
Event Date 04/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an exactamix eva total parenteral nutrition (tpn) bag had a pinhole along the side seam near the top of the bag which resulted in a leak.This event was discovered during setup and before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed and a small tear in the top left side seam of the bag was observed.A functional leak test identified a leak in the top left side seam of the bag.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
see h10
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7492358
MDR Text Key107966620
Report Number1416980-2018-02668
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477268
UDI-Public(01)00085412477268
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2020
Device Catalogue NumberH938742
Device Lot Number1231275
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/21/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-