This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
|
Product analysis: analysis could not confirm the reported issue.It was noted that there were errors in the external pulse generator (epg) logs, the main printed circuit board (pcb) was electrically out of specification.The upper case was cracked at the bosses, the main seal was missing, and the display wire was pinched.Additionally, the lower case and three main pcb screws were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device was re-calibrated and functionally tested.The device passed all final testing.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
|