• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/17/2018
Event Type  malfunction  
Manufacturer Narrative
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) was freezing or hanging up after it was turned on.The epg is expected to be returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: analysis could not confirm the reported issue.It was noted that there were errors in the external pulse generator (epg) logs, the main printed circuit board (pcb) was electrically out of specification.The upper case was cracked at the bosses, the main seal was missing, and the display wire was pinched.Additionally, the lower case and three main pcb screws were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device was re-calibrated and functionally tested.The device passed all final testing.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7495130
MDR Text Key107695567
Report Number3004593495-2018-00476
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-