• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB CRYSTALENS ACCOMMODATING IOL; LENS, INTRAOCULAR, ACCOMMODATIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB CRYSTALENS ACCOMMODATING IOL; LENS, INTRAOCULAR, ACCOMMODATIVE Back to Search Results
Model Number AO1UV
Device Problems Difficult to Advance (2920); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/10/2018
Event Type  Injury  
Manufacturer Narrative
Investigation of this event is in progress.A follow-up report will be submitted upon completion of the investigation.
 
Event Description
It was reported that the back of the lens got stuck in the ci-28 injector during the procedure.The incision was enlarged but no sutures were required.Additional information has been requested, but no additional information has been received.
 
Manufacturer Narrative
The lens was not returned for evaluation; therefore, a product evaluation could not be performed.The device history record (dhr) was reviewed and there were no discrepancies or deviations found.Based on available information, a root cause for the reported event could not be conclusively determined; however, it is possible that user related factors (such as loading or handling technique) and/or procedural factors (such as lens and injector interaction) may have caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYSTALENS ACCOMMODATING IOL
Type of Device
LENS, INTRAOCULAR, ACCOMMODATIVE
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
21 north park place blvd.
clearwater FL 33759
Manufacturer Contact
tes proud
1400 north goodman street
rochester, NY 14609
5853388549
MDR Report Key7496846
MDR Text Key107685394
Report Number0001313525-2018-00099
Device Sequence Number1
Product Code NAA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date10/31/2020
Device Model NumberAO1UV
Device Catalogue NumberAO1UV-2150
Device Lot Number7568211
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-