BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/11/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold¿ lite with capio slim device was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio cage failed to catch the needle.After several attempts, the carrier cracked.The procedure was completed with another uphold¿ lite with capio slim device.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
A visual examination of the returned device revealed that the capio slim head was separated at the top allowing the carrier to protrude through the head, preventing the dart from being caught in the cage.All rivets were attached.The carrier was not broken.Analysis revealed no damage to the uphold lite mesh assembly.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold lite with capio slim device was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio cage failed to catch the needle.After several attempts, the carrier cracked.The procedure was completed with another uphold lite with capio slim device.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|