Catalog Number 9001C0212
Device Problem Hole In Material (1293)
Patient Problem No Patient Involvement (2645)
Event Date 04/18/2018
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) of rebt2285 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the device presented a hole.It was further reported that the cover presented a hole as well.The device was not used on a patient.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a hole present in the probe cover is confirmed and appears to be use related.One site rite needle guide kit packaging with lot: rebt2285 and ref: 9001c0212 and one probe cover was returned for investigation.The probe cover was covered in gel.The probe cover was filled with water to determine the hole location.Two leaks were observed approximately 1.3 inches from the closed end of the cover.The leak locations were approximately 1 inch apart.A scratch in the cover was observed near the two leak locations.Microscopic observation revealed a slit in the plastic cover at the leak sites.The scratch in the cover and the two slits appear to have been caused by contact with a sharp instrument; therefore, the complaint is confirmed, use related.A lot history review (lhr) of rebt2285 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the device presented a hole.It was further reported that the cover presented a hole as well.The device was not used on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE ULTRASOUND SYSTEM NEEDLE GUIDE KIT (18G) (POLYETHYLENE) (48IN)
Type of Device
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key7498918
MDR Text Key107828545
Report Number3006260740-2018-00937
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00801741037818
UDI-Public(01)00801741037818
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K042445
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number9001C0212
Device Lot NumberREBT2285
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Is the Reporter a Health Professional? Yes
Event Location Hospital
Date Manufacturer Received06/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-