The recipient was reimplanted with another cochlear device.The external visual inspection revealed the electrode was severed along the lead near the array prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The residual gas analysis result was above the limit, indicating a non-hermetic device.Due to the presence of moisture getter, no signs of moisture were observed.The failure of this device is attributed to a short from a power node to the electrode ground case at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on the electrode ground node inside the analog chip integrated circuit.This ultimately caused the device to cease functioning.A corrective action was implemented.This is the final report.
|