• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number SE-05-040-120-6F
Device Problems Difficult or Delayed Positioning (1157); Physical Resistance (2578)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation summary: a visual, dimensional and functional inspection was performed on the returned device.The deployment difficulty and difficulty with the thumbslide were not able to be confirmed as the stent had already been deployed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other similar incidents.The investigation was unable to determine a cause for the reported deployment difficulty and difficulty advancing the thumbslide.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was performed to treat a lesion with moderate calcification.Pre-dilatation was performed with an unspecified armada 18 balloon catheter.A 6f sheath was selected for use.A 5.0 x 40 mm supera veritas self-expanding stent was advanced to the target lesion.Although there were difficulties rotating the thumbslide, the stent was successfully deployed in the target lesion.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7501092
MDR Text Key107963816
Report Number2024168-2018-03432
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Catalogue NumberSE-05-040-120-6F
Device Lot Number7090561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-