• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL CAVITRON TOUCH ULTRASONIC SCALING SYSTEM; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL CAVITRON TOUCH ULTRASONIC SCALING SYSTEM; SCALER, ULTRASONIC Back to Search Results
Model Number G1000
Device Problems Restricted Flow rate (1248); Overheating of Device (1437); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g1000 scaler, the water flow could not be regulated, the water was hot, and the handpiece was getting hot; no injury resulted.
 
Manufacturer Narrative
Evaluation found water filter filled with debris, causing restricted water flow.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON TOUCH ULTRASONIC SCALING SYSTEM
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7502753
MDR Text Key107984057
Report Number2424472-2018-00067
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K150535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG1000
Device Catalogue NumberG1000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Was the Report Sent to FDA? No
Date Manufacturer Received06/01/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-