BAXTER HEALTHCARE - SINGAPORE FLOGARD; PUMP, INFUSION Back to Search Results
Catalog Number 2M8063F
Device Problem Low Battery (2584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/16/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a flogard had a low battery.The process step when the event occurred is unknown.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was serviced on-site by a field service technician.Visual inspection, functional testing, and a review of the alarm log were performed.The low battery alarm was identified during functional testing and the review of the alarm log.The cause of the condition was determined to be a depleted battery.To correct the condition, the battery was replaced.The device was serviced to meet functional specification.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOGARD
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
singapore
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7507574
MDR Text Key108019217
Report Number1416980-2018-02794
Device Sequence Number1
Product Code FRN
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2M8063F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-