The conmed affiliate reported that the mw-319, tran histology needle, device tip fell off when used.The incident reportedly took place on (b)(6) 2018; however, the type of procedure being performed was not disclosed.The reporter has stated that there was no injury to patient or user.There was no report of delay of procedure or cancellation of the procedure.Additionally, further information has been requested; however, no response has been received to date.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
The alleged needle, mw-319, was not returned to conmed for evaluation.This complaint of "needle falling off" was unable to be verified and a root cause cannot be determined.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 5 complaints, regarding 6 devices, for this device family and failure mode.During this same time frame 50,848 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.0001.Per the instructions for use, the user is advised the following: precautions: - never force the instrument into the channel - make sure the needle tip is within the metal hub prior to use - retract the needle tip into the metal hub prior to removal from the channel this issue will continue to be monitored through the complaint system to assure patient safety.
|